- Simultaneous detection of up to 3 different toxic gases.
- Up to 3-year gas sensor warranty.
- Outstanding durability: Passes 1.5 m drop testing, equivalent to IP66/68 protection rating
- Bluetooth® equipped
- Inherently safe explosion-proof construction allows use in explosion-proof areas.
- Continuous operating time: Approx. 60 hours
|
Sampling method |
Suction type/Double type |
|
Principle of detection |
Potentiostatic Electrolysis Method |
|
Detection range |
Depend on target gas. |
|
Alarm setpoints |
Depend on target gas. |
|
Type of alarm |
Gas alarm / fault alarm |
|
Display of alarm |
Lamp flashing / continuous modulating buzzer sounding / gas concentration reading blinking |
|
Explosion protected construction |
Intrinsic safety |
|
Explosion-proof class |
ATEX : Ⅱ 1 G Ex ia ⅡC T4 Ga |
|
Ingress proof rating |
IP66/68 |
|
Approvals |
Japan Ex, ATEX, IECEx |
|
CE marking |
EMC / ATEX / RoHS / RE |
|
Power source |
Lithium ion battery unit [BUL-9000] or |
|
Continuous operation |
Approx. 60 hours [Lithium ion battery unit] |
|
Dimensions |
Approx. 158 (W) × 85 (H) × 132 (D) mm (excluding projections) |
|
Weight |
Approx. 1.1 kg (includes battery and battery unit) |
|
Operating temperature range |
-20 ℃ to +50 ℃ (no sudden changes) |
|
Operating humidity range |
20 to 90 %RH (no condensation) |
|
Operating pressure range |
80 kPa to 120 kPa (80 to 110 kPa for explosion-proof range) |
|
Target gas |
CL2, HCL, HF, NH3, NO, NO2, O3, PH3, SiH4 |